8 Maintaining the MMI
To Change the Liner and O-Ring on the MMI
136 Maintaining Your GC
Clean the inlet if there is visible or suspected contamination. See “To Clean the
Multimode Inlet” on page 139.
Clean O-ring residue from sealing surface.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
7 Slide a new O-ring onto the replacement liner.
8 Install the liner to the inlet, pushing it all the way in until the liner contacts the seal.
9 Line up the tab on the bottom of the insert assembly with the slot on the inlet body and
push down to connect. Slide the locking tab to the back.
10 Turn on the inlet. Allow the inlet and column to purge with carrier gas for 15 minutes before
heating the inlet or the column oven.
11 Bakeout contaminants. See “To Bakeout Contaminants from the MMI” on page 141.
12 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
Restriction tests, and will automatically reset the maintenance counters.
13 Select Finished, then select OK to exit the GC maintenance wizard.
14 Restore the analytical method.