9 Maintaining the PTV Inlet
To Replace the Septumless Head PTFE Ferrule on the PTV Inlet
Maintaining Your GC 153
12 Observe the inlet Total Flow. Tighten the guide cap until the Total Flow stops decreasing
(typically near 8 mL/min).
13 Remove the syringe from the inlet and press .
14 Restore the analytical method.
15 Perform an inlet Leak & Restriction test and reset the maintenance counters.