EasyManua.ls Logo

Agilent Technologies 8890 - To Check FID Performance

Agilent Technologies 8890
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14 Maintaining the ECD
To Bakeout the ECD
Maintaining Your GC 239
8 Allow thermal cleaning to continue for several hours and then cool the system to normal
operating temperatures. The figure below shows detector output during a typical cleaning
cycle.
9 Check the ECD output value on the control table. It should be lower than the first reading. If
it is not, contact your Agilent service representative.
10 Reinstall the column.
11 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
12 Restore the analytical method.

Table of Contents

Other manuals for Agilent Technologies 8890

Related product manuals