15 Maintaining the FPD+
To Change the FPD+ Wavelength Filter
248 Maintaining Your GC
5 To prevent light from damaging the PMT, cap the end or place it face down.
6 Place a clean cloth under the filter housing to catch the filter.
For phosphorus filter, use the sharpened wooden tip of a toothpick or cotton swab to
dislodge the filter from the housing.
For sulfur filter (shown below), use the wooden tip of the cotton swab to remove the
filter spacer. Then dislodge the filter from the housing.