5 Maintaining the Split/Splitless Inlet
To Install a Capillary Column with the Split/Splitless Inlet
Maintaining Your GC 57
16 Install the column into the detector.
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
To Install a Capillary Column in the ECD
To Install a Capillary Column to the FPD+
17 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
18 After the column is installed at both inlet and detector, establish a flow of carrier gas and
purge as recommended by the column manufacturer.
19 Restore the analytical method.
For FID or FPD+, immediately turn off the flame.
For NPD, immediately turn off the bead.
20 After the GC becomes ready, wait 10 minutes then ignite the detector flame or bead.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If they are hot,
wear heat-resistant gloves to protect your hands.
21 Allow the oven, inlet, and detector to equilibrate at operating temperature, then retighten
the fittings.
22 Perform an inlet Leak & Restriction test and reset the maintenance counters.