5 Maintaining the Split/Splitless Inlet
To Clean the Split/Splitless Inlet
70 Maintaining Your GC
13 Perform an inlet Leak & Restriction test and reset the maintenance counters.
14 Bakeout contaminants. See “To Bakeout Contaminants from the Split/Splitless Inlet” on
page 71.
15 On the GC touchscreen, select Maintenance > Inlets, then touch Reset Counter.
16 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
17 Restore the analytical method.