6 Maintaining the Purged Packed Inlet
To Change the Glass Liner on the Purged Packed Inlet
Maintaining Your GC 89
6 Wearing gloves, grasp the flared end (top) of the replacement glass liner with tweezers and
install it in the inlet. If the glass liner does not seat properly because a capillary column is
installed, remove the column, install the glass liner, and replace the column. See “To Install
a Capillary Column with the Purged Packed Inlet” on page 78.
7 Install the top insert assembly and hand-tighten firmly.
8 Configure the new liner.
9 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
Restriction tests, and will automatically reset the maintenance counters.
10 Select Finished, then select OK to exit the GC maintenance wizard.
11 Restore the analytical method.