6 Maintaining the Purged Packed Inlet
To Clean the Purged Packed Inlet
Maintaining Your GC 91
15 Install the septum and septum nut. See “To Change the Septum on the Purged Packed
Inlet” on page 82.
16 Attach the column. See “To Install a Capillary Column with the Purged Packed Inlet” on
page 78.
17 Set the septum and glass liner counters.
18 Perform the inlet Leak & Restriction test and reset the maintenance counters.
19 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
20 Restore the analytical method.