7 Replacement Procedures
Replacement procedure for Micro GC sample line
Agilent 990 Micro GC User Manual 99
8 Use a 3/16 inch wrench to screw the new sample line nut to the inlet manifold.
9 Install the inlet manifold using two Torx T-10 screws.
10 Use a 3/16 inch wrench to screw the new sample line nut to the front of the channel.
11 Tuck the sample line into the GC cabinet so that the line does not interfere with replacement of
the front cover.
12 Replace the front cover.