ESG-D Series Option H98 Signal Generators In–Channel, Base Station and Mobile Receiver Testing
Reconciling New Firmware Revisions with Earlier Revisions
Manual Supplement 1-7
Reconciling New Firmware Revisions
with Earlier Revisions
Remote and front-panel operation of firmware revision B.02.30 and greater is quite
different from earlier firmware revisions. This section documents the remote operation
commands from earlier firmware revisions that are still valid in revision B.02.30.
To determine the firmware revision of your instrument, press
Utility > Instrument Info/Help
Mode
> Diagnostic Info. To locate the firmware revision information on the display, see the
following illustration.
Remote Operation
The following SCPI commands were used in previous versions of firmware. Although they
are different than the commands for the current firmware, they can still be used.
System Frame Number Query
This command query retrieves the current value of System Frame Number from the
wideband CDMA hardware.
[:SOURce]:RADio:WCDMa[:BBG]:SFN?
Wideband CDMA Off On
This command sets the wideband CDMA modulation operating state to on or off. The
command query will retrieve wideband CDMA operating state information from the signal
generator.
[:SOURce]:RADio:WCDMa[:BBG][:STATe] ON|OFF|1|0
[:SOURce]:RADio:WCDMa[:BBG][:STATe]?
Firmware Revision Number