EasyManua.ls Logo

Agilent Technologies G8610A - Main Settings

Agilent Technologies G8610A
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent Technologies Helium Leak Detector
104
DRAFT 12
/11/17
6.2 Main Settings
To open this screen select Menu tab > Setting > Main Settings.
With the Settings Menu open, select the Main Settings Menu as shown below:
Figure 6-1 Settings: Main Settings
Buttons and Fields
Date & Time Display and configure the current date and time.
Units Select leak rate units:
❑ atm-cc/sec
❑ mbar-l/sec
❑ Torr-l/sec
❑ Pa-m3/sec.,
❑ Cubic- ft./yr.
Select pressure units:
❑ Torr
❑ Mbar
❑ Pascal

Table of Contents