Agilent Technologies Helium Leak Detector
118
DRAFT 12
/11/17
7.4.3 Manual Tuning
Press Manual Tuning and the Manual Spectrometer Tuning screen appears
(
Figure 7-7).
Use this screen to view and manually adjust the parameters of the ion source in the
spectrometer for a maximum helium signal.
Figure 7-7 Manual Tuning
Emission Current Displays the value of the existing Emission current of the spectrom-
eter Ion Source and allows for emission current adjustment. The
emission current parameter tunes the leak detector, maximizing the
leak detector output for a peak helium signal.
Typical Emission current values range from 0.6 mA to 1.2 mA.
Ion Voltage Displays the current ion voltage value of the spectrometer Ion
Source. The ion voltage parameter tunes the leak detector, maxi
-
mizing the leak detector output for a peak helium signal.
Typical ion voltage values range from 205 VDC to 240 VDC. The
ion voltage value is automatically adjusted and set during the full
calibration routine.
Increments/
the selected value
decrements
Select the
filament for use.
Select a value
to change it.
Accesses the
keypad to set Gain.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G8610A and is the answer not in the manual?

Agilent Technologies G8610A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG8610A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals