NOTE The password can also be read through the RS-232 port in the
event of a lost password. See
Section A “Communications
Protocol” on page 139.
Lock Vent Control Disables the VENT button. This option prevents accidental system
venting during test while under remote or PLC control. The Vent
Lock feature is not active when the Auto Sequencer is on.
Lock All Control Disables the Test, Vent and Zero buttons. This option prevents acci-
dental system venting, testing, system hold and Zeroing during test
while under remote or PLC control.
6.3 Wireless
❑ Information relating to wireless communications.
MAC Address Two MAC addresses appear on this screen:
❑ The MAC address at the top of the screen is the leak detector
MAC address.
❑ The MAC address above the hand held button is the MAC
address of the hand held device.
SYNC Channel For informational and factory debug purposes only.
SS ID Displays the SS ID number, factory set to 086, for informational and
factory debug purposes only.
Channel Click the button and a keypad appears to configure the channel
(1-19) that the unit uses to communicate with the wireless remote.
Up to 10 wireless remotes and leak detectors can be
communicating in a common area at a time.