EasyManua.ls Logo

Agilent Technologies G8610A

Agilent Technologies G8610A
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent Technologies Helium Leak Detector
DRAFT 12/11/17
39
background and shows the time until the calibration begins. Should you prefer not
to wait, a Cancel button permits the calibration request to be canceled.
NOTE The firmware initially starts at a large time interval, but
recalculates the required time as the pumping continues. The
new time is reflected on the timer.
NOTE Test Configuration Menu > Applications Setup > High
Sensitivity automates and walks you through the E-10 External
Calibration procedure.
Figure 1-16 Helium Background Timer
NOTE When calibration is complete, the leak detector reverts to
Fine
Test mode.

Table of Contents

Related product manuals