EasyManua.ls Logo

Agilent Technologies Infinity II 1290

Agilent Technologies Infinity II 1290
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
214 Agilent 1290 Infinity II Flexible Pump User Manual
11 LAN Configuration
Manual Configuration
6 Change the IP address (in this example 134.40.27.99) and type “/” to list
current settings.
Figure 41 Telnet - Current settings in "Using Stored" mode
information about the LAN interface
MAC address, initialization mode
Initialization mode is Using Stored
active TCP/IP settings
TCP/IP status - here ready
connected to PC with controller software (e.g. Agilent
ChemStation), here not connected
Figure 42 Telnet - Change IP settings
change of IP setting to
Initialization mode is Using Stored
active TCP/IP settings
stored TCP/IP settings in non-volatile memory
connected to PC with controller software (e.g. Agilent
ChemStation), here not connected

Table of Contents

Related product manuals