22 Agilent SPS 4 Autosampler User’s Guide
2 Introduction
Specifications
Environmental
Your Agilent SPS 4 is designed for indoor use only. It is suitable for
Installation Category II and Pollution Degree 2. The site should be selected
to avoid dusty or corrosive atmospheres.
Power
The power cable that is included with the Agilent SPS 4 has an AC
adapter. It is rated at the following.
Table 1 Suitable environmental conditions for the Agilent SPS 4
Environmental Condition Specification
Temperature Range: Non-Operating -20°C60°C, 2080 % RH
-4
°F140°F, 2080 % RH
Temperature Range: Operating 10°C35°C, 2080 % RH
50°F95°F, 2080 % RH
Operational Altitude 02000 m
06562 feet
Table 2 Input/output rating of the AC adapter for the Agilent SPS 4
Power Specification
Power Supply: Input: 100240V, 47/63Hz, 1.5A.
Output: 24 VDC, 2.5 A
Autosampler Input Power: Input: 24 VDC, 2.5 A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies SPS 4 and is the answer not in the manual?

Agilent Technologies SPS 4 Specifications

General IconGeneral
BrandAgilent Technologies
ModelSPS 4
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals