EasyManuals Logo

Agilent Technologies SPS 4 User Manual

Agilent Technologies SPS 4
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #79 background imageLoading...
Page #79 background image
Running the Calibration Wizard 7
Agilent SPS 4 Autosampler User’s Guide 79
10 When all four points have been captured, a confirmation message is
displayed asking if you want to write the new calibration values.
Click Yes to write the calibration values.
The autosampler is adjusted using the new calibration value.
Click No if you want to retry.
The calibration process is restarted.
11 After calibration has been successfully completed, check the positioning
of the sample probe by clicking on each corner position button, and
driving the Z- Axis position down. Confirm that the sample probe is
positioned in the center of each well.
Error List
On running the calibration wizard software, you may encounter any error
listed below:
•Serial Port Error
Example: Autosampler did not respond.
A serial port error occurs if the software cannot communicate with the
autosampler. Check that the power is on, and no error lights are
illuminated. Also confirm the USB connection and the COM port setting
in the instrument control software. If the problem still remains
unresolved, contact your Agilent customer support representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies SPS 4 and is the answer not in the manual?

Agilent Technologies SPS 4 Specifications

General IconGeneral
BrandAgilent Technologies
ModelSPS 4
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals