EasyManua.ls Logo

Agilent Technologies SPS 4 - Installing the Sample Racks

Agilent Technologies SPS 4
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40 Agilent SPS 4 Autosampler User’s Guide
3 Installation
Figure 11 Installing the standards rack
Installing the Sample Racks
The sample racks are located on both (left and right) sides of the
standards rack by mating the rack’s feet with the holes in the rack
location mat. The rack location mat has maximum four sample rack
positions
two rack positions on the either side of the standards rack
position. The following rack options are supported.
21 position rack for 30 mm OD tubes
24 position rack for 25 mm OD tubes
40 position rack for 20 mm OD tubes
60 position rack for 17 mm OD tubes
90 position rack for 13 mm OD tubes
To install the sample racks:
1 Locate the holes for the sample racks in the rack location mat.
2 Align the feet from each sample rack with the holes in the rack location
mat. Press down on the rack to ensure it is correctly located and the
bottom of the sample rack is sitting flush with the rack location mat.

Table of Contents

Related product manuals