Installation 3
Agilent SPS 4 Autosampler User’s Guide 37
4 Position the rack location mat over the spill tray and align the dovetail
feature in the mat with the dovetail feature in the left end frame before
pressing down firmly.
5 Check that the rack location mat fits firmly in place on the spill tray by
trying to move it from side to side and front to back. There should not
be any movement when fitted correctly.
6 If the rack location mat does not fit correctly even after completing the
steps above, contact your local Agilent office or representative.
Installing the Wash Reservoir
Figure 8 Single port fixed wash reservoir and kit
The fixed wash reservoir is supplied as a kit with the inlet and outlet
nipples fitted along with some lengths of inlet (2.5 mm ID) tubing and
outlet (5.0 mm ID) tubing. For the single port fixed wash reservoir, rinse
solution flows up through the smaller wash chamber and overflows into
the larger drain chamber to waste. For the dual port fixed wash reservoir,
there are two wash chambers and a common drain chamber.
The fixed wash reservoir is mounted at the rear of the rack location mat
(see Figure 6).
To install the fixed wash reservoir:
1 Put the base of the fixed wash reservoir into the locating hole and twist
it clockwise 60 degrees so it locks into position. The nipples on the
fixed wash reservoir should be facing toward the rear of the

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies SPS 4 and is the answer not in the manual?

Agilent Technologies SPS 4 Specifications

General IconGeneral
BrandAgilent Technologies
ModelSPS 4
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals