74 Agilent SPS 4 Autosampler User’s Guide
7 Running the Calibration Wizard
Performing User Calibration
To perform user calibration:
1 Ensure the autosampler power is on.
The sample racks fitted should be empty and not contain any sample
vials. If a microtiter plate Adapter is fitted, it should be loaded with
four 96 well microtiter plates.
2 Start the Calibration Wizard software. The Calibration Wizard window
opens as shown below.
3 Select the appropriate COM port from the drop- down list, and then
click Connect in the Com Port field.
If you chose the wrong one, an error message appears. If the
appropriate COM port number is not listed, check the autosampler
power, the USB connection, and the COM port setting in the instrument
control software.
4 If the calibration wizard can connect to your autosampler successfully,
the field under the Com Port field shows the information on the rack
layout specific to your autosampler. Ensure that the sample rack types
displayed on the screen correspond to those loaded into their respective
positions on the rack location mat.
NOTE
Close the instrument software before launching the Calibration Wizard.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies SPS 4 and is the answer not in the manual?

Agilent Technologies SPS 4 Specifications

General IconGeneral
BrandAgilent Technologies
ModelSPS 4
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals