Troubleshooting 6
Agilent SPS 4 Autosampler User’s Guide 67
0 R 004RAM test error Reinitialize the SPS 4. If
error persists, contact
your authorized Agilent
representative.
0 R 0 R 5 Axis initialization error Reinitialize the SPS 4. If
error persists, contact
your authorized Agilent
representative.
0 RR0 6 X-axis position error. The X-axis motor
was commanded to move, but did not
reach the desired position.
Refer to “Probing
Problems” on Page 69
for details.
0 RRR7 Theta-axis position error. The
Theta-axis motor was commanded to
move, but did not reach the desired
position.
Refer to “Probing
Problems” on Page 69
for details.
R 0008Z-axis position error. The Z-axis motor
was commanded to move, but did not
reach the desired position.
Refer to “Probing
Problems” on Page 69
for details.
R 00R 9 EEPROM error Reinitialize the SPS 4. If
error persists, contact
your authorized Agilent
representative.
R 0 R 0 10 Mat checksum error Reinitialize the SPS 4. If
error persists, contact
your authorized Agilent
representative.
R 0 RR11 X-axis movement detected without the
motor being commanded to move.
Refer to “Probing
Problems” on Page 69
for details.
RR0 0 12 Theta-axis movement detected without
the motor being commanded to move.
Refer to “Probing
Problems” on Page 69
for details.
Table 7 Error Indicator Codes
E1 E2 E3 E4 Error
Code
Error Description Action