Running the Calibration Wizard 7
Agilent SPS 4 Autosampler User’s Guide 77
8 When point “Rear Right” is captured, the sample probe moves to point
“Front Left” above the front left sample well in the sample rack or
microtiter plate in that position, and the indicator next to the button
labeled “Front Left” changes to orange.
To adjust the probe position in point “Front Left”:
a Adjust the sample probe height to the top of the sample well using
the Up/Down buttons.
b Adjust the XY position to the middle of the well using the arrow
buttons.
c When the position of the probe is centered in the sample well, click
Capture.
The indicator next to the “Front Left” button changes from orange to
green.