Revision H, 11/10 VK 7025 Dissolution Apparatus Page 79
P/N 70-9033 Operator’s Manual Operation
Varian, Inc.
Step 6. Continue the dissolution test according to the prescribed method. Two
minutes after either the cannulas return to the home position or the final
keystroke, the Test in Progress screen displays.
Paddle Over Disk
Step 1. From the Start Method screen, select INSTANT START.
Step 2. If paddle spin is enabled, press STOP. If paddle spin is disabled, skip this
step and continue to step 3.
Note
To quit the test from the Test in Progress screen, press ESC once
to return to the Method Status screen and a second time to quit
the test.
Note
Before starting the test, ensure paddle spin is enabled or disabled
(see “Paddle Spin” on page 46) and final temp is enabled or
disabled (see “Sample Vessel Temp” on page 88) depending on
your configuration.
Paddle Spin Final Temp
After selecting INSTANT START, the following
information displays:
enabled disabled or enabled
METHOD STARTUP STATUS / MANUAL DISK DROP / PRESS
STOP TO STOP PADDLES / PRESS RUN TO CONTINUE
disabled disabled or enabled METHOD STARTUP STATUS / MANUAL DISK DROP / PRESS
RUN TO CONTINUE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Varian VK 7025 and is the answer not in the manual?

Agilent Technologies Varian VK 7025 Specifications

General IconGeneral
BrandAgilent Technologies
ModelVarian VK 7025
CategoryMedical Equipment
LanguageEnglish

Related product manuals