Page 82 VK 7025 Dissolution Apparatus Revision H, 11/10
Operation Operator’s Manual P/N 70-9033
Varian, Inc.
If you press STOP / PAUSE, the Method Paused screen displays. The time paused and
the current elapsed time display. Select EXIT to stop the current method.
ABORTING
METHOD STAND BY displays and the screen returns to the Start Method screen. To continue
the test, select CONTINUE. The Method Status screen displays.
Delay Start
If you are using paddles, DDMs, and an autosampler, you can program a delayed start
based on a specific date and time. If you do not select paddles and DDM in your method,
this option is not available. To set the delayed start, complete the following steps:
Step 1. From the Start Method screen, select DELAY START. The Time Delay
Start Setup screen displays the current date and time below the RPM
and water bath temperature. If temp display is enabled (see “Temp
Display” on page 46), the vessel temperatures / cannula positions
display as well.
Step 2. The cursor flashes under
ENTER DATE. Enter the date in the appropriate
format and press ENTER.
ENTER TIME displays.
Step 3. Enter the time in 24-hour format (for example, 3:00 pm is entered as
15:00:00) and press ENTER. The Time Delay Start Status screen
displays the time until the test starts. Once the test begins, the Method
Status screen displays (see page 81).
Note
For this option, if you enter a value outside of the acceptable
range, an error message displays indicating appropriate values.
Note
To abort the test from either the Time Delay Start Status screen
or the Method Status screen, press and hold ESC. Once the
method begins, press STOP / PAUSE or ESC and select EXIT.
The Start Method screen displays.