VS Series Helium Mass Spectrometer Leak Detectors
DRAFT 4/23/15
3-23
3.1.8.3.1 Cal Leak Expiration Set-Up
Press the button next to the Cal Leak Expiration text on the Internal Calibrated Leak
screen.
The Cal Leak Expiration Set-Up screen appears (Figure 3-18).
The Cal Leak Expiration is usually one year from leak installation. When the
calibrated leak expires, a message appears on the Home screen.
Figure 3-18 Cal Leak Expiration Set-Up
To use thi s s c r e en:
1. Press the button next to Year or Month and a keypad appears to set the value.
2. Press Done for each item configured.
3. Press the button next to Day, use the up and down arrows to set the value and press
Done.
4. Press OK to exit the screen.
012345678910
-
00
CAL LEAK EXPIRATION SET-UP
DONECANCEL
OK
YEAR
MONTH
2006
AUG
DAY
02
Accesses the
Ye a r k e y p a d .
Activates the field.
Use the arrows to
set the value
Accesses the
Day keypad.