VS Series Helium Mass Spectrometer Leak Detectors
DRAFT 4/23/15
A-1
Appendix A. Communications Protocol
This appendix defines the technical specifications for the RS-232 protocol used with the
leak detector and describes the customer-configured inputs/outputs.
A.1 Protocol (RS-232)
RS-232 ports operate at 9600 baud, 8 bits, no parity, and one stop bit.
All characters transmitted to the leak detector are echoed by the leak detector. Commands,
inquiries, and strings of both commands and inquiries must be terminated by a carriage
return <CR>; the carriage return is echoed as a space character.
The maximum length of the input is 80 characters; if no carriage return has been received
before the 80th character, execution of the command string begins anyway.
Words that begin with:
? are inquiries for the controlling device to determine the current state or value of a
leak detector parameter.
PUT- are commands for the controlling device to set the current state or value of a
volatile leak detector parameter.
INIT- are commands for the controlling device to set non-volatile leak detector
parameters.
Other commands do not require a parameter and do not begin with any special characters.
Successful inquiries respond with data, followed by a space, then: ok<CR><LF>, as
specified in the tables of this appendix. Unsuccessful inquiries respond with the offending
inquiry followed by a space, then: #?<CR><LF> (line feed, <LF>).
Inquiries, parameters, and commands can be concatenated. Each word or numeric
parameter is followed by one or more spaces. The string is terminated by a <CR>, which
causes execution to begin.
Successful strings respond with the specified data for the input inquiries in the order the
inquiries were issued, followed by: ok<CR><LF>.
Unsuccessful strings respond with the first offending command followed by a space,
then: #?<CR><LF>. All commands and inquiries following the failed word are ignored;
all parameters are discarded.
Table A-1 on page A-5, Table A-2 on page A-8, Table A-3 on page A-11, and Table A-4 on
page A-11 outline the available control and query commands.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies VS Series and is the answer not in the manual?

Agilent Technologies VS Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelVS Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals