EasyManuals Logo
Home>Agilent Technologies>Security Sensors>VS Series

Agilent Technologies VS Series User Manual

Agilent Technologies VS Series
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
VS Series Helium Mass Spectrometer Leak Detectors
DRAFT 4/23/15
A-9
INIT-ION Sets the ion voltage of the ion source. Preceded by a three-digit number
(volts) in the range 220 to 280.
INIT-LANGUAGE Selects the language to be shown on the display. Preceded by 0-6 where
(0=English, 1=French, 2=German, 3=Korean, 4=Spanish, 5=Japanese,
6=Chinese).
INIT-LINEAR Sets the leak rate analog output voltage to Linear. Not preceded by a value.
INIT-LKEXPIRE Sets the date on which the internal calibrated leak expires. Preceded by the
date.
Example: mm dd yyyy INIT-LKEXPIRE
INIT-LEAKTEMP Sets the internal calibrated leak temperature (printed on the leak label)
that the factory calibration was performed at. Preceded by 12.0 to 50.0
(degrees C).
Example: 23.5 INIT-LEAKTEMP
INIT-1LOG Sets the leak rate analog output voltage to logarithmic at 1 volt per decade.
Not preceded by a value.
INIT-ML-XFER Sets the value of the midstage-flow mode crossover pressure in Torr.
Preceded by X.XE-XX. The range is 1.0E-03 to 1.0E-01 Torr. The default is
1.0E-01 (100 mTorr).
INIT-PASSWORD Sets the system password. Preceded by a 1 to 6 digit password and
followed by a digit indicating the number of digits in the password.
Example: 123123 6 INIT-PASSWORD.
Password is read back using the ?SYSPWD command.
X INIT-PWDLEVEL X is the password level from 1 to 3.
INIT-QUICK-CAL Sets which type of calibration will be run if the CALIBRATE command is
initiated. Preceded by 0 or 1, 0 for FULL cal, 1 for FAST cal.
INIT-RANGESTOP Sets the value of the most sensitive range exponent. Preceded by XX.
INIT-1REJECT Sets the leak rate value for Reject Set Point #1. Preceded by a two-digit
leak rate mantissa with a decimal point after the first digit followed by E-,
and then by a two-digit leak range exponent, the helium leak rate number
is in atm cc/sec.
Example 3.4E-08 INIT-1REJECT
INIT-2REJECT Same as INIT-1REJECT, but for Reject Set Point #2.
INIT-3REJECT Same as INIT-1REJECT, but for Reject Set Point #3.
Table A-2 Non-Volatile Operating Parameters (Continued)
Command Parameter

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies VS Series and is the answer not in the manual?

Agilent Technologies VS Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelVS Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals