VS Series Helium Mass Spectrometer Leak Detectors
DRAFT 4/23/15
1-29
Pressure Indication Test port a n d s p ectrometer p r e s s ure with alp h a n u meric indi c a t i o n on
the LCD display.
Spectrometer Sensitivity optimized design, header mounted ion source with dual
thoria-coated iridium filament, a high sensitivity, wide dynamic range
preamplifier, and vacuum gauge.
Automatic Valve Sequencing Microprocessor-controlled valve operation provides automatic system
cycling.
Safety Interlocks A valve system design that completely protects the high vacuum
system under test against pressure surge and power interruption.
Optional
Wireless Remote Control This provides:
RF communication between the VS leak detector (with base sta-
tion or host) and the Wireless Remote unit from a distance up to
100 m (300 ft).
The ability to enter the VS leak detector unit into Test and Hold
modes from the Wireless Remote unit's keypad.
The ability to read the VS leak detectors leak rate in a linear and
logarithmic bar graph format with a numeric range indicator and
System status in alphanumeric format on the Wireless Remote
unit's LCD display.
Configuration up to 10 different VS leak detector unit’s for wireless
control by toggling through a MAC address list using one Wireless
Remote unit
Discrete I/O Control This provides 12 outputs and 8 inputs lines, optically isolated, to
control the VS leak detector using an external PLC via a 25-pin DB25
female connector and a user-provided parallel cable.
Table 1-4 VS Series Features (Continued)
Feature Description

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies VS Series and is the answer not in the manual?

Agilent Technologies VS Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelVS Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals