PN 903-8797, Rev A
Special Conditions for Safe Use
During use, the Cooler Housing section of the Measuring Cell Assembly can
functionally be purged with a dry gas such as air or nitrogen. The pressure of
this gas at the upstream end of the breathing and draining device is maximum
7 kPag (1 PSIG).
The sample gas must be above the upper explosion limit and may have a
maximum pressure of 138 BAR, at a temperature of +40 °C (+104 °F).
Ex d certied conduit or cable seals that comply with the ameproof enclosure
cable entry sealing requirements of the local authority must be installed at
the entries to the Electronics Enclosure. All unused cable entry ports must be
plugged with a certied Ex d plug.
x | 241CE II Hydrocarbon Dewpoint Analyzer