EasyManuals Logo

Ametek 241CE II User Manual

Ametek 241CE II
171 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
PN 903-8797, Rev A
To set the purge ow rate:
1. First, determine the steady-state ow through the Flow Restrictors in the
Filterblock.
2. Close the pressure regulator.
3. Adjust the pressure regulator for the lowest ow rate (turn counter-clock-
wise as far as possible) before reading the ow rate on the ow meter.
4. While the analyzer is in its Purge cycle (Purging/Hold Stage), adjust the
pressure regulator until the ow has increased by the desired purge ow
rate, approximately 1.0–5.0 Nl/minute (2.1–10.6 SCFH)
[Total ow = steady state + purge ow].
Normal Operation
Once the analyzer is set up and started, the measuring cycle, the automatic
scrolling of results on the User Interface, and the presentation of measurement
results on the analog output channels is continuous. Typically, the only atten-
tion required will be when it is necessary to change an operating parameter or
to clear any current Warning or Fault alarms. This can be done from the
Stat\History menu (“Clear Alarm? prompt).
2-40 | 241CE II Hydrocarbon Dewpoint Analyzer

Table of Contents

Other manuals for Ametek 241CE II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 241CE II and is the answer not in the manual?

Ametek 241CE II Specifications

General IconGeneral
BrandAmetek
Model241CE II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals