PN 903-8797, Rev A
To set the purge ow rate:
1. First, determine the steady-state ow through the Flow Restrictors in the
Filterblock.
2. Close the pressure regulator.
3. Adjust the pressure regulator for the lowest ow rate (turn counter-clock-
wise as far as possible) before reading the ow rate on the ow meter.
4. While the analyzer is in its Purge cycle (Purging/Hold Stage), adjust the
pressure regulator until the ow has increased by the desired purge ow
rate, approximately 1.0–5.0 Nl/minute (2.1–10.6 SCFH)
[Total ow = steady state + purge ow].
Normal Operation
Once the analyzer is set up and started, the measuring cycle, the automatic
scrolling of results on the User Interface, and the presentation of measurement
results on the analog output channels is continuous. Typically, the only atten-
tion required will be when it is necessary to change an operating parameter or
to clear any current Warning or Fault alarms. This can be done from the
Stat\History menu (“Clear Alarm? prompt).
2-40 | 241CE II Hydrocarbon Dewpoint Analyzer