EasyManuals Logo

Ametek 241CE II User Manual

Ametek 241CE II
171 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
PN 903-8797, Rev A
Adjusting the Sample Gas Flow Rate
The gas sample ow rate needs to be properly adjusted for your application.
During each measurement cycle, the solenoid is closed. The main purge ow
through the solenoid is also shut o. During the Purging/Hold Stage, the
solenoid opens and the sample gas ows through the sample system and
Measuring Cell. This stage allows previously condensed liquids to evaporate,
while obtaining a new representative gas sample. During this stage, the gas
purge ow rate must be high enough (e.g.,1.0–5.0 Nl/minute or 2.1–10.6 SCFH)
to completely purge the sample system and Measuring Cell within the allotted
Purge time.
To set the ow rate, use the pressure regulator on the backpan (Figure 2-1). If
this adjustment has been done properly during installation, no further adjust-
ment should be necessary.
If you measure the ow rate at the Vent Line, you will nd that it is sub-
stantially higher during the Purging/Hold Stage because the Purging/
Hold Stage ow rate (1.0–5.0 Nl/minute (2.1–10.6 SCFH) is added to
the trickle purge ow rate (Bypass Valve is open).
Because high ow rates result in increased pressure drops through the
Membrane Filters, do not set the purge ow rate greater than 5.0 Nl/minute
(10.6SCFH). The higher pressure tends to force liquids through the mem-
branes. The purge ow rate setting is especially important if hydrocarbon
and/or water droplets are present in the gas stream. Hydrocarbons have low
surface tension and can be forced through Membrane Filters more easily than
high surface tension liquids such as glycol, alcohols, and water. Liquid carry-
over through the lter can result in measurement problems. It becomes dif-
cult – or even impossible – to detect condensing hydrocarbons when there is
already hydrocarbon liquid present.
The sample ow rate and the Purge time can be easily adjusted. Therefore, it is
relatively easy to obtain a reasonable balance between the two variables. The
rst thing to determine is whether to operate the analyzer with a bypass ow
(or trickle purge”) during the measuring cycle. Although the solenoid shuts o
the main purge ow during the measure cycle, opening the Bypass Valve will
allow a small ow rate during measurement (approximately one-tenth of the
purge cycle ow). This bypass ow tends to improve accuracy and response
time because there is always a small, but constant ow of sample gas owing
through the sample system and Measuring Cell.
Installation and Start-Up | 2-37

Table of Contents

Other manuals for Ametek 241CE II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 241CE II and is the answer not in the manual?

Ametek 241CE II Specifications

General IconGeneral
BrandAmetek
Model241CE II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals