EasyManua.ls Logo

Ametek 241CE II

Ametek 241CE II
171 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PN 903-8797, Rev A
Filterblock Maintenance
The Membrane Filters in the Filterblock Assembly should be inspected periodi-
cally for signs of contamination or damage.
Clean the Filterblock:
If alarms that indicate the optical paths show signs of contamination are
displayed, such as:
- f H2O Optics (SCode 32)
- f HC Optics (SCode 64)
If there is evidence of baseline deviations (see “Checking and Recording
the Baseline Settings” in the Diagnostic Flowcharts section of this chapter).
Each time you clean the Measuring Cell.
If the internal parts are severely contaminated, more frequent cleaning
may be required.
Replace parts in the Filterblock:
If, while cleaning the Filterblock, the Membrane Filters are found to be
damaged (are creased, have holes, or are transparent) or severely contami-
nated (with liquids, oils, or waxes).
See “Spare Part Lists”
in this chapter for ordering
numbers.
RUN Mode
Stat\HistorySCode
4-22 | 241CE II Hydrocarbon Dewpoint Analyzer

Table of Contents

Other manuals for Ametek 241CE II

Related product manuals