PN 903-8797, Rev A
11. Using the alignment pins as guides, make sure that the electrical con-
nectors are properly aligned and then push the Cooler Housing directly
toward the Electronics Enclosure. While holding the Measuring Cell/Heat-
sink Assembly rmly in place with one hand, fully thread in one (1) of the
upper screws. Reinstall the remaining three (3) screws and tighten them
in a cross-pattern. The torque specication is 4.0 Nm, ±0.7 Nm (36 in.-lb,
±9 in.-lb).
12. Reconnect the tubing to the inlet and outlet ttings on the Measuring Cell
and hand-tighten the nut for each tting. Using two wrenches, use one
to rotate the nut to the original position (an increase in resistance will be
encountered at the original position), then tighten it slightly. Repeat for
the other tting.
13. Reconnect the Purge Air Connection tubing to the Cooler Housing.
14. Perform a leak check on the sample system ttings that were disconnect-
ed.
15. Using a soft, nonabrasive cloth, gently clean the joining areas (ame path)
of the Electronics Enclosure and its door. Replace and tighten the M10
screws to 9.0 Nm, ±0.7 Nm (80 in.-lb, ±9 in.-lb).
16. Restore AC power to the analyzer. After start-up, the analyzer will immedi-
ately begin operation. The Run Time is reset to zero (‘0’).
17. Reset the Mirror Low setpoint temperature (MTLo) and the purging time
(Purge) to their normal (original) operating values.
The results obtained by the rst few measuring cycles are usually
slightly too low (invalid) due to contamination. This can be caused by
the residual water remaining from the cleaning (after maintenance)
and/or exposure of the sample system to the atmosphere – i.e., diluted
with residual air after installation or maintenance.
Depending on the degree of contamination and the purging duration,
the number of measuring cycles required to obtain valid results can
vary. Typically, ve measuring cycles will be sucient to remove the
water or air contaminants after the sample system has been exposed
to water or the atmosphere.
See “Sample System
Leak Check” in Chapter 2.
Maintenance and Troubleshooting | 4-21

Table of Contents

Other manuals for Ametek 241CE II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 241CE II and is the answer not in the manual?

Ametek 241CE II Specifications

General IconGeneral
BrandAmetek
Model241CE II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals