Maintenance
Generally, there is limited maintenance required to ensure the analyzer re-
mains operating at peak eciency, other than that described in this section.
This maintenance, which includes parts that require replacement and cleaning,
is intended to ensure continued and proper operation of the analyzer.
The following sections discuss parts that can be replaced in the analyzer”
Performing maintenance on the analyzer requires working from the
User Interface. Familiarize yourself with how to work from the User
Interface before working on the analyzer (see Chapter 3).
Cleaning or Replacing the Sample and Vent Lines
The Sample and Vent Lines should be inspected periodically, and any time
problems associated with the analyzers measurements may be in question.
Generally, it is not necessary to clean the Sample and Vent Lines. However, if
there are liquids present in the lines, purge the lines with a dry gas or instru-
ment air.
Determining whether cleaning or replacing the line is necessary will depend
on the design of the Sample Line, and the availability of replacement tubing
and/or a cleaning medium. Example, replacing an unheated Sample Line may
be more practical than cleaning it. Cleaning the Sample Line with steam may
work if the contaminant is water soluble.
Whether cleaning or replacing the line, always purge the line with a dry gas
before reconnecting it to the analyzer.
Generally, it is necessary to replace the sample lines if the Sample or Vent Lines
are found to be sagging, have been bent, or show damage to the outer skin.
For tools and supplies required to replace the Sample/Vent Lines, see Tools,
Equipment, and Supplies Required” in Chapter 2. If cleaning is required, use
Isopropanol or another similar non-abrasive cleaning agent.
Maintenance and Troubleshooting | 4-11

Table of Contents

Other manuals for Ametek 241CE II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 241CE II and is the answer not in the manual?

Ametek 241CE II Specifications

General IconGeneral
BrandAmetek
Model241CE II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals