PN 903-8797, Rev A
Clearing (Resetting) Alarms
Active alarms can be cleared (reset) from the Stat\History menu. Clearing the
alarms removes any active alarms from the User Interface display and stores
them (up to a maximum of nine) in the History Buer (^SC or ^XSC), also in
the Stat\History menu. A password is not required to perform this command.
To clear alarms:
1. From the RUN menu, press to view the Stat\History menu. The rst
menu item you will see is the En\Dis Alrm? prompt.
2. Press once to view the Clear Alarm? prompt. While this prompt is on
the User Interface, press the + button to clear the alarms; the message
Cleared will appear on the User Interface.
If you do not wish to clear the alarms, press the button; the message
NotCleared will appear on the screen. Or, you can press to scroll past
the “Clear Alarm? prompt and leave the alarms in their current state.
Enabling/Disabling the O-Spec Alarm
The O-Spec Alarm can be enabled (turned on) or disabled (turned o) from
the Stat\History menu. When the O-Spec Alarm is enabled, LED 1 is O;
when the O-Spec Alarm is disabled, LED 1 is On to alert the operator to this
condition. A password is not required to perform this command.
To enable or disable the O-Spec Alarm:
1. From the RUN menu, press to view the Stat\History menu. The rst
sub-menu item you will see is the En\Dis Alrm? prompt.
2. If the O-Spec Alarm is already in the current state you require, press to
scroll past the En\Dis Alrm? prompt and leave this parameter in its cur-
rent state.
To enable the O-Spec Alarm if it is currently disabled, press the + button.
LED 1 will turn O to indicate the O-Spec Alarm is enabled.
To disable the O-Spec Alarm if it is currently enabled, press the button.
LED 1 will turn On to indicate the O-Spec Alarm is disabled.
Maintenance and Troubleshooting | 4-41

Table of Contents

Other manuals for Ametek 241CE II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 241CE II and is the answer not in the manual?

Ametek 241CE II Specifications

General IconGeneral
BrandAmetek
Model241CE II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals