EasyManuals Logo

Ametek 933S User Manual

Ametek 933S
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #79 background imageLoading...
Page #79 background image
Installation and Start-Up | 3-33
h. From the Output Setup dialog box, verify the current and voltage
output ranges and concentration alarm setpoints are configured
correctly for the application:
Below each configured channel (Assignment), confirm that the
Iout/Vout Low Scale and Iout/Vout High Scale are the correct
minimum and maximum ranges required.
To clamp the V/I output result at 4 mA, enable V/I Low Limit
4 mA; the output will not go below 4 mA.
To confirm a change, click OK then Apply; otherwise, click
Cancel.
Verify that all concentration alarms are configured correctly for
the application. The alarm setpoints are set at the factory using
application-specific data. If changing setpoints, see “Output
Setup” in Chapter 4.
NOTE
Figure 3-16.
Output Setup dialog
box.
i. From the Alarm Setting and Stream Switch Parameters dialog box
(Figure 3-17), verify that the Concentration Alarms (1 and 2) are
configured correctly for the application.
If you need to alter the ranges, refer to “Alarm Setting and Stream
Switch Parameters” in Chapter 4.
The alarm setpoints (On-threshold/Off-threshold) are set at the
factory using application-specific data.
After configuring the Concentration Alarm setpoints, click OK then
Apply. Save the changes to EEPROM.
NOTE
Setup (tab)Alarm/
Stream Switch
Setup (tab)
Output Setup

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 933S and is the answer not in the manual?

Ametek 933S Specifications

General IconGeneral
BrandAmetek
Model933S
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals