Section 5 GPIB
5-12
6) AUX auxiliary output
Selects the signal output from the rear AUX terminal.
Command : AUX n
Query : AUX?
Response : AUX n
<Program data>
Value of n Value set
0 .......................... (initial value)
1 .......................... Go/No-Go
2 .......................... Count End
3 .......................... Level Det
4 .......................... Int Gate
5 .......................... Restart
6 .......................... Acquisition
Go/Nogo : Outputs the decision result of the template function.
When High, the measurement frequency is within the setting range.
When Low, the measurement frequency is outside of the setting range.
Outputs High when the template function is not selected.
Coun End : A Low pulse is output each time frequency measurement ends.
Level Det : Outputs a counter internal detection signal during burst signal measurement.
Int Gate : Outputs an internal gate signal used in the frequency count.
Restart : Outputs a Low pulse when the Restart command is executed.
Acquisition : Outputs Low during acquisition operation.
(2) B
1) BST burst measurement
Specifies whether to perform burst measurement or CW measurement.
Command : BST n
Query : BST?
Response : BST n
<Program data>
Value of n Value set
0 .......................... BURST OFF : CW measurement (initial value)
1 .......................... BURST ON : Burst measurement