5-19
5.4 Device Message List
2) MBWDT measurement data (burst width)
Outputs burst width when there is one measurement result read function during burst measurement.
Query : MBWDT?
Response : n
<Response data>
Value of n ........... Output is units of time (NS)
During CW measurement (when burst is Off), 0 NS is returned.
3) MBPRD measurement data (burst period)
Outputs burst period when there is one measurement result read function during burst measurement.
Query : MBPRD?
Response : n
<Response data>
Value of n ........... Output in units of time (NS)
During CW measurement (when burst is Off), 0 NS is returned.
4) MCW measurement data (continuous wave)
Outputs a frequency measurement value when there is one measurement result read function during CW measure-
ment.
Query : MCW?
Response : n
<Response data>
Value of n ........... Output in frequency (HZ) units.
Returns 0 HZ in other cases (during burst measurement).
5) MOFS measurement data (offset frequency)
Outputs the +/–offset calculation result and ppm calculation result with a single measurement read function.
Query : MOFS?
Response : n
<Response data>
• +Offset or –Offset :
Value of n ....... Output in frequency (HZ) units.
• ppm :
Value of n ....... Output in deviation (ppm) units.
Returns 0 HZ in other cases.