5-21
8) MDS measurement data (frequency from the data storage memory)
Reads the data traced in the internal memory.
100 pieces of data are output, starting with the oldest one (r
1
).
Query : MDS?
Response : r
1
r
2
:
T
100
(10) O
1) OFS offset
Adds the frequency value obtained in advance to the frequency measurement result and calculates the subtraction
or deviations
Command : OFS n (, s)
Query : OFS?
Response : OFS n
<Program data>
Value of n Value set
0 .......................... Off (initial value)
1 .......................... +OFFSET On
2 .......................... –OFFSET On
3 .......................... ppm
Value of s Value set
0 .......................... Offset value of the value set by the command OFSFRQ. (default value)
1 .......................... Offset value of the previous measurement value.
(The value set by OFSFRQ is overwritten.)
2) OFSDT offset data
Selects whether to switch the offset value update mode On or Off. When the update mode is On, the previous
measurement values are successively updated as offset values.
Command : OFSDT n
Query : OFSDT?
Response : OFSDT n
<Program data>
Value of n Value set
0 .......................... Update mode Off (initial value).
1 .......................... Update mode On
5.4 Device Message List