5-25
(12) S
1) SH sampling hold
Starts and stops frequency measurement.
Command : SH n
Query : SH?
Response : SH n
<Program data>
Value of n Set value
0 .......................... Sampling (Initial value)
1 .......................... Hold
Note :
When the unit is in the Hold state (SH1), it is restarted with either *TRG or GET (address commands).
2) SMP sample rate
Sets the sampling rate (pause time).
Command : SMP n
Query : SMP?
Response : SMP n
<Program data>
Value of n Set value
0 .......................... 1 ms
1 .......................... 2 ms
2 .......................... 5 ms
3 .......................... 10 ms
4 .......................... 20 ms
5 .......................... 50 ms
6 .......................... 100 ms (Initial value)
7 .......................... 200 ms
8 .......................... 500 ms
9 .......................... 1 s
10 ........................ 2 s
11 ........................ 5 s
12 ........................ 10 s
5.4 Device Message List