Section 5 GPIB
5-24
D6 : Indicates the exponent for numeric data.
E+0=10
0
, E+3+10
3
, E+6+10
6
, E+9+10
9
D7 : Dummy data
LF ^ EOI : TRM0 (Initial value)
CR LF ^ EOI : TRM1
(11) R
1) REF reference frequency
Selects whether to switch the reference signal Auto or Internal.
Command : REF n
Query : REF?
Response : REF n
<Program data>
Value of n Set value
0 .......................... Auto (Initial value)
1 .......................... Internal
2) RES frequency resolution
Sets the frequency measurement resolution.
Command : RES n
Query : RES?
Response : RES n
<Program data>
Value of n Set value
0 .......................... 1 mHz
1 .......................... 10 mHz
2 .......................... 100 mHz
3 .......................... 1 Hz
4 .......................... 10 Hz
5 .......................... 100 Hz (Initial value)
6 .......................... 1 kHz
7 .......................... 10 kHz
8 .......................... 100 kHz
9 .......................... 1 MHz
3) RTM return to measure
Displays the measurement screen.
Command : RTM