5-27
(13) T
1) TRG trigger mode
Selects the trigger source.
Command : TRG n
Query : TRG?
Response : TRG n
<Program data>
0 .......................... INT (Initial value)
1 .......................... EXT
2 .......................... LINE
2) TRGDLY trigger delay
Sets the trigger delay value.
Command : TRGDLY n
Query : TRGDLY?
Response : TRGDLY n
<Program data>
Value of n
0,
20 10
–9
to 100 10
–3
(sec) ............. Suffix : NS, US, MS, S, N, U, M (Unit sec)
Note that you should set n to a value between 20 ns to 320 ns in increments of 20 ns, 320 ns to 1 µs in increments
of 40 ns, or 1 µs to 100 ms with two significant digits. If you make a setting outside of those ranges, it will be
rounded down. In addition, setting a value of 20 ns or lower sets the delay to Off.
3) TRGPL trigger edge polarity
Sets the trigger detection polarity.
Command : TRGPL n
Query : TRGPL?
Response : TRGPL n
<Program data>
Value of n Set value
0 .......................... positive (Initial value)
1 .......................... negative
5.4 Device Message List