Section 5 GPIB
5-28
4) TRM terminator
Selects the terminator when sending response data.
Program message : TRM n
<Program data>
Value of n Set value
0 .......................... LF (Initial value)
1 .......................... CR LF
5) TRS transient mode
Sets the high-speed sampling function ON/OFF.
Command : TRS n
Query : TRS?
Response : TRS n
<Program data>
Value of n Set value
0 .......................... Off (Initial value)
1 .......................... On
Note :
High-speed sample measurement is started with either *TRG or GET (address commands).
6) TRSOFS transient offset
Outputs the standard frequency fo to the input frequency calculation using high-speed sampling. (see MTRS
message)
Query : TRSOFS?
Response : n
<Response data>
Value of n ........... Output in frequency (HZ) units.
Sending back 0 Hz when selects the input2.
7) TRSSMP transient sample point
Sets the number of points for measuring using the high-speed sampring function.
Command : TRSSMP n
Query : TRSSMP?
Response : TRSSMP n
<Program data>
Value of n
100, 200, 500, 1000, 2000 (Initial value)