PU-4010 4-2
Turn off the power.
• In the case of AC-power operation, detach the cord of the AC adapter from the instrument.
Remove the test strip holder from the test strip holder base.
• Slide the test strip holder approx. 20 mm to the right to unlock. Lift up the holder without
hitting the holder base.
Wash the test strip holder and test strip
holder base.
• Clean the holder and holder base with a
neutral detergent and carefully wash adhered
urine off with running water. Use a soft piece
of cloth to wipe water off the holder and
holder base and dry completely.
Wipe dirt off the grooves of the guide rails.
• Use a soft piece of cloth or cotton swab dampened with neutral detergent diluted with
water to wipe dirt off the guide rails. Use a soft piece of cloth to wipe water off and dry the
guide rails completely.
5
NOTE: Be sure to turn off the power. Cleaning the guide rails with the power on may
damage the instrument.
6
2. Lift up the holder.
1. Slide the holder 20 mm to the right.
Test strip holder
Lift up the holder without hitting
the holder base.
Test strip holder base
Side view of
the test strip holder
Test strip
holder base
Test strip holder
7
8