How to Use - Setting the Incubator
20
1-5 O
2
Setting
This item allows to preset the concentration of O
2
gas to be used.
Press and hold [SET] for 1 second or longer to make the temperature indication flash. Then,
press SILENCE twice to make O
2
concentration indication flash. When it is flashing, press
SET to switch the indication to O
2
setting indication .
To change the current setting, press SET again to make the temperature setting indication
flash, and then press UP or DOWN to select the desired value. After changing the setting,
press SET to complete the setting.
To return to the previous layer of the setting item, press SILENCE.
Example of O
2
setting indication (Numeric value: Shown for reference)
1-6 Flowchart of the Setting Modes
Normal display
1 second
Temp. blinks
Channel Select
Change with
Change with
Change with
Change with
Change with
Change with
Temp. Setting
Temp. Change
CO
2
blinks
CO
2
setting
CO
2
change
Over Temp. Setting
Over Temp. Change
O
2
blinks
O
2
setting
O
2
change
Under Temp. Setting
Under Temp. Change