How to Use - Fine Adjustment of the Incubator
26
1-18 O
2
Alarm Setting
O
2
alarm setting allows to set a tolerance for alarm generation, relative to the O
2
concentration
preset through O
2
setting. If 1.0 is set for example, the tolerance will be ±1.0%.
Press and hold SILENCE for 1 second or longer to display temperature fine-adjustment
mode indication .
When in this display, press UP twice or DOWN four times to display O
2
fine-adjustment
mode indication .
Then, press SET to switch the indication to O
2
SPAN setting indication .
Moreover, when in this display, press UP twice or DOWN twice to display O
2
alarm
setting indication .
To change the current setting, press SET again to make the O
2
alarm setting indication flash,
and then press UP or DOWN to select the desired O
2
alarm value. After changing the
setting, press SET to complete the setting.
To return to the previous layer of the setting item, press SILENCE.
Example of O
2
alarm setting indication (Numeric value: Shown for reference)
1-19 O
2
Alarm Delay Time Setting
O
2
alarm delay time setting allows to set delay time to alert.
Alarm will be triggered when current O
2
concentration on the display deviates more than the
preset at O
2
alarm setting after the time set in this menu has passed. The unit of the numeric
value to be set is "minute".
Press and hold SILENCE for 1 second or longer to display temperature fine-adjustment
mode indication .
When in this display, press UP twice or DOWN four times to display O
2
fine-adjustment
mode indication .
Then, press SET to switch the indication to O
2
SPAN setting indication .
Moreover, when in this display, press UP three times or DOWN once to display O
2
alarm
delay time setting indication .
To change the current setting, press SET again to make the O
2
alarm delay time setting
indication flash, and then press UP or DOWN to select the desired O
2
alarm delay time
value. After changing the setting, press SET to complete the setting.
To return to the previous layer of the setting item, press SILENCE.
Example of O
2
alarm delay time setting indication (Numeric value: Shown for reference)