Ch. 11. Tutorial: AC Mode in Air, ES Sec. 11.2. Loading the Cantilever and Sample
9.
Set aside the cantilever holder:
Set the cantilever holder on its handles
with the cantilever facing up so that you
do not crush the cantilever.
10.
Locate your cantilever holder:
Identify the appropriate cantilever
holder. This tutorial requires the
standard Gas cantilever holder, Asylum
Part# 901.758.
Note To learn more about cantilever holders
for the Environmental Scanner, please refer to
Chapter 12 on page 141.
11.
Prepare cantilever mounting workspace:
You will need the following items:
a) Tweezer, curved (Asylum Part
#290.102)
b) 300 Philips Screwdriver (Asylum Part
#290.106)
c) Box of AC160TS cantilevers
d) Gas cantilever holder
A low power binocular dissection
stereoscope with light source can be
useful for some of the following steps.
Cleaning the tweezer tips with alcohol
improves the handling of the cantilevers.