Chapter 4 – Functions of the instrument
UM-22100H-U User manual ATEQ D520 Page 91/122
3.4. INFINITE FILL
This function is validated when an optional shut off valve is installed in the instrument.
This valve allows the stopping of the flow of air (or of gas) out of the cycle.
The infinite fill allows the opening of this shut off valve, in other words to create a
constant flow at the test pressure, allowing the part to be checked and to adjust and fine
tune the assembly, etc.
From the main menu, position
the cursor in front of the
CYCLE SPE title and confirm
with the ENTER key.
MAIN
MENU
ACTIVE PROG. : 001
PARAMETERS
CYCLE SPE: Disabled
Next, place the cursor in front
of Infinite fill and validate with
the ENTER key.
SPE
CYCLE
Disabled
Regulator adjust
Infinite fill
The display confirms that the
special cycle is selected.
MENU
PRINCIPAL
RUN PROG. : 001
PARAMETERS
SPE CYCLE :Inf Fill
Press on the START key to
start the special cycle.
CYCLE/P
r:
001
PRESS = 20 mbar
FILL
Adjust the value of the test
pressure with the regulator
handle.
_
+
CYCLE/P
r:
001
PRESS = 20 mbar
FILL
To stop the cycle, press on
the RESET key.
CYCLE/P
r:
001
PRESS = 20 mbar
READY