EasyManuals Logo

ATEQ D520 User Manual

ATEQ D520
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
Chapter 1 – Installation of the instrument
UM-22100H-U User manual ATEQ D520 Page 17/122
2.2. CONNECTOR DETAILS
2.2.1. Electrical connectors
The ATEQ D520 operates under a power of 24V DC with either :
9 the transformer delivered with the instrument,
9 or through the network cable when the instrument is a slave.
2.2.1. 1) J1 connector (Output codes / Analogue outputs /
Temperature sensors)
1 2 3 4 5 6 7 8 9 10111213141516
Pin 1 COMMON (outputs 1, 2, 3) + 24 V DC
Pin 2 Output n°1, open collector
Pin 3 Output n°2, open collector
Pin 4 Output n°3, open collector
Pin 5 COMMON (outputs 4, 5, 6) + 24 V DC
Pin 6 Output n°4, open collector
Pin 7 Output n°5, open collector
Pin 8 Output n°6, open collector
OUTPUT CODES
24V DC 100mA Max
outputs
Pin 9 12V DC Sensor power supply (2 mA max)
Pin 10 0V Sensor power supply
Pin 11 Sensor input n°1
Pin 12 Sensor input n°2
TEMPERATURE
SENSOR
Pin 13 Analogue output n°1
Pin 14 COMMUN (analogue output 1)
Pin 15 Analogue output n°2
Pin 16 COMMUN (analogue output 2)
ANALOGUE
OUTPUTS
24 V DC
8
7
6
5
4
3
2
1
0,1 A max
Charge / Load
Obligatory
diode for an
inductive load.
2.2.1. 2) J2 connector
Telephone network socket. Not used.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ATEQ D520 and is the answer not in the manual?

ATEQ D520 Specifications

General IconGeneral
BrandATEQ
ModelD520
CategoryTest Equipment
LanguageEnglish

Related product manuals