3.5. PIEZO RESET (AUTO ZERO)
Enables to carry out of a forced auto zero (reset) of the piezo sensor.
In the main menu, place the
cursor in front of the SPE.
CYCLE option and confirm
with the ENTER key.
MAIN
MENU
RUN PROG. : 001
PARAMETERS
SPE CYCLE: Disabled
Itif
Next, place the cursor in front
of Auto-zero Piezo and
confirm with the ENTER key.
SPE
CYCLE
Pre-regul. adjust.
Infinite fill
Auto-zero Piezo
The display confirms that the
special cycle is selected.
MAIN
MENU
RUN PROG. : 001
PARAMETERS
SPE CYCLE: AZ-Piezo
Press on the START key to
start an auto zero cycle.
RUN
/P
r :
001
PRESS = 355.5 mbar
AUTO_ZERO
Once the auto zero is carried
out, the cycle will stop.
RUN
/P
r :
001
PRESS = 355.5 mbar
READY
Note: the auto zero is an automatic event which is carried out for each cycle, it enables
the initialization of the pressure sensors to the atmospheric pressure.